How can I mount a pulse oximeter on the surface of PCB without putting too much stress?

The PCB has a pulse oximeter mounted on the surface, and it's tolerance is 0.6mm. The problem is the pulse oximeter needs to touch the bottom case at all times. How to best mount the PCB in the case without putting too much stress on the pulse oximeter? Will screws with flexible washers be all right? Sadly separating pulse oximeter from main PCB is not an option.

Put a machined guard around the pulse oximeter to keep the surface from it at about 0.2-0.3 (if it is and SMD part soldering height tolerance is about 0.2mm) and then simply screw down the resulting assembly.

Otherwise use weak springs to push the p.o. against the case, need to know the allowed stress on the body for that.

Read More: Medical Electrics Manufacturing

#PCB Assembly

https://www.youtu.be/kX4gCs9szeA?si=_o1w0lNIu4HwRX7N

Oliver Smith

Oliver Smith

Oliver is an experienced electronics engineer skilled in PCB design, analog circuits, embedded systems, and prototyping. His deep knowledge spans schematic capture, firmware coding, simulation, layout, testing, and troubleshooting. Oliver excels at taking projects from concept to mass production using his electrical design talents and mechanical aptitude.
Oliver Smith

Oliver Smith

Oliver is an experienced electronics engineer skilled in PCB design, analog circuits, embedded systems, and prototyping. His deep knowledge spans schematic capture, firmware coding, simulation, layout, testing, and troubleshooting. Oliver excels at taking projects from concept to mass production using his electrical design talents and mechanical aptitude.

What Others Are Asking

How can I determine BGA land pad diameter for given ball diameter?

I’m working on a project which requires the use of a CSP package. The product’s datasheet provides ball pitch and ball diameter but nothing about the preferred land pad diameter.
How can the pad diameter be worked out from these figures assuming an NSMD land?

Can leaded BGA chips be reflow soldered with lead-free process?

I know that using leaded process for lead-free BGA chips is not acceptable, because the solder balls of the BGA will not melt and the joint will be unreliable. But now BGA chip is only available in lead-free balls. is it ok to solder leaded BGA chips with a lead-free (thus, higher temperature) process?

Are SMT components bad for high voltage applications?

Many assembly factories are asking for SMT jobs, while I think through hole would be a better option for a high voltage application. Before the high voltage project is started, we need to make a call on SMT or Through hole parts. Is there a study on this?

Read Detailed Advice From Blog Articles

Scroll to Top